MediSense PCx manual Review/Change Test Setup for Patient Test

Models: PCx

1 151
Download 151 pages 28.1 Kb
Page 73
Image 73

Review/Change Test Setup for Patient Test:

The options set with this menu will apply to all Patient Tests.

The Patient ID option uses a patient identification number to perform tests. Patient ID uses the following options:

Disable, Require or make Optional the entry of a Patient ID.

Require a minimum and maximum number of digits for the length of the Patient ID.

Allow any combination of five common barcode formats used in the facility.

I2 of 5 (with or without check digit)

Code 39 (with or without check digit, alphanumeric or ASCII format)

Code 128 Codabar Code 93

Truncate (ignore) leading and/or trailing barcode digits.

Comment Code option allows the following to be selected:

Disable the prompt for a Comment Code.

Prompt for a Comment Code with the option of not requiring it.

Require a Comment Code for every test result out-of-range and/or in-range of the action range. (See Test Type setup options).

The Memory Full option will notify the operator that the patient test memory has reached its limit (4000 tests that have not been uploaded). Memory Full provides the following options:

Lockout additional tests until the data is uploaded.

Prompt the operator to upload data before overwriting old test results with new ones.

Automatically overwrite oldest test results with new ones.

8-13

Questions? Refer to Page 10-13 For Customer Care Information.

Page 73
Image 73
MediSense PCx manual Review/Change Test Setup for Patient Test