What You See | What You Do | Comments | |
on the Display | |||
|
| ||
|
|
|
8.Gently invert the required control solution bottle
9.Wait for the monitor to analyze the control solution and display the test result.
10.Note the test result and whether it falls within the acceptable range.
(2.7 mmol/L)
The monitor beeps when the sample is accepted. The Sample Accepted screen will be displayed. Recap the control solution bottle tightly.
For a detailed description of the test strip target area, refer to page
The test starts automatically as the sample is accepted.
The monitor counts down 20 seconds then displays the test result.
Control test results can appear as numeric or PASS/FAIL, depending on how the test is setup.
The display also shows the control level, the control range, date and time.