What You See on the Display
What You Do | Comments |
|
|
5.Scan or manually enter the linearity kit lot number via the keypad, then press Enter. If the monitor prompts for a level to be entered, skip to step 7.
If the linearity kit lot number that was entered is different from the previous one, the monitor will prompt the operator to do one of two actions: either
6.The operator can perform one of the following actions:
Press 1 – ReEnter
Kit Lot.
Press 2 – Replace Panel.
If 1 is pressed, the monitor will return to the previous screen so the operator can scan or enter the linearity kit lot number.
The monitor will prompt the operator to select a level.
7.Enter the number of the level of the next test to be run.
8.If the operator presses 6 for a new panel, the monitor will again prompt the operator to select a level.
The monitor will prompt the operator to scan or manually enter the test strip lot number.
There can be up to 4 replicates per level. The number in the parenthesis indicates the number of replicates already run for that level. When a level is full, the menu item (1- Level 1) will no longer appear.