MediSense PCx manual Linearity Test

Models: PCx

1 151
Download 151 pages 28.1 Kb
Page 57
Image 57

Linearity Test

7

 

 

The Linearity Test menu enables the operator to verify method linearity using a Linearity Assessment Kit.*

Please use the following procedures to perform a Linearity Test.

If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.

What You See

What You Do

Comments

on the Display

 

 

 

 

 

1.Press On/Off to turn on monitor.

The Abbott logo screen will appear for a few seconds, followed by the product name, the software version, then the Test Mode menu.

2.Press Menu.

3.Press 3 – Linearity Test.

4.Press Scan or manually enter the Operator ID via the keypad, then press Enter.

The Menu Mode menu will appear.

The monitor will prompt for the Operator ID to be scanned or entered.

The monitor will next prompt for the Linearity Kit lot number to be scanned or entered.

*It is outside the intended use of the PCx/QCM system to perform a test in linearity mode on a PCx instrument and change the test result record to a patient result in QCM Data Management System. Linearity test results outside the test strip assay range should only be used to set linearity performance on the PCx instrument. The performance of the test strips has not been verified outside the assay range listed on the test strip instructions for use.

7-1

Questions? Refer to Page 10-13 For Customer Care Information.