What You See on the Display
What You Do | Comments |
|
|
4.Press Scan to scan the Patient ID barcode, or manually enter the Patient ID via the keypad, then press Enter.
After entering the Patient ID, the test strip lot screen appears next.
5.Press Scan to scan the test strip barcode, or manually enter* the test strip lot number via the keypad, then press
Enter.
6.The operator can select one of the following:
Press 1 – Capillary or Arterial.
Press 2 – Venous.
7.Open the foil test strip packet at the notch and tear up or down to remove the test strip.
After the test strip lot number has been entered, the Sample Type screen will appear if this setting has been enabled. If it does not appear, skip down to step 7.
Note: Use VENOUS MODE only for venous samples that have been in capped tubes and have not been mixed with air. Use CAPILLARY or ARTERIAL MODE for all other samples.
* Manual entry of strip lots with
Questions? Refer to Page