MediSense PCx manual Review/Change Test Setup for Security

Models: PCx

1 151
Download 151 pages 28.1 Kb
Page 85
Image 85

Review/Change Test Setup for Security:

Refer to the beginning of the Review or Change Setup section for the step-by-step instructions to access the appropriate setup menu.

If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.

What You See on the Display

What You Do

Comments

 

 

1.Press 1 – Test Setup.

2.Press 4 – Security.

3.The administrator will select one of the following options:

Press 1 – Next, to continue to Security Operator ID Not Certified screen and skip forward to step 10.

Press 2 – Change, to change the QC Interval Controls Expired option.

The display shows the Test Setup menu.

The display shows the Security QC Interval menu.

The display shows the options for the QC Interval Controls Expired menu. The options are to Allow Test, even if the QC interval has expired; Warn the operator that even though the QC interval has expired, the test may continue; or Lockout the system until the required control tests are performed.

8-25

Questions? Refer to Page 10-13 For Customer Care Information.

Page 85
Image 85
MediSense PCx manual Review/Change Test Setup for Security