5.Where a TSE_DOLBYE (RVF) test sequence is used, make sure that the option to use blanking and ancillary data From Source Image (RVF) is selected in the Video: Generator page of the Engineering Settings control window. This setting is essential for correct operation.

Where the PCM-only (YUV) test sequence is being used, use the Audio: Generator page of the Engineering Settings control window to generate an audio tone on channel 1 is set for 8kHz, -18dB, and present only on frame 1 of the sequence.

Also ensure that the option to use blanking and ancillary data From Internal Generator is set on the Video: Generator page of the Engineering Settings.

These settings are essential for correct operation.

Page 112

OmniTek XR User Guide, Software Release 2.3

Page 112
Image 112
OmniTek 2.3 manual OmniTek XR User Guide, Software Release