113
Triggering Examples
Single-Machine Trigger Examples
This sequence specification causes the analyzer to trigger when the
data does not occur in 16 ms or less. If it does occur within 16 ms, the
sequence restarts. Specifications of this type are useful in finding
intermittent problems. You can set up and run the trace, then cycle the
system through temperature and voltage variations, using automatic
equipment if necessary. The failure will be captured and saved for later
review.
Triggering when Data not Returned