Retrieving and Restoring Data Using the LAN
Retrieving and Restoring Data Using the LAN
To copy raw measurement data
1Set up the measurement you want to make, and run the analyzer
to acquire data.
For more information on setting up measurements, see the
Logic Analyzer section of this book. You can set up and run the
measurement from the analyzers XWindow interface or front panel, or
by programming the analyzer.
2Access the logic analyzers file system.
Refer to the chapter "Accessing the Logic Analyzer File System".
3Copy the data.raw file from the \slot_a directory for analyzer
data, or \slot_b directory for oscilloscope data.
Raw measurement data files are binary format files that can be
transferred to your computer and then reloaded into the logic analyzer
later. Because they are binary files, if you are using ftp to upload or
download them, set the type to binary by typing "bin" at the ftp prompt.
The data.raw file is overwritten each time new data is acquired.