520
Retrieving and Restoring Data Using the LAN
Retrieving and Restoring Data Using the LAN
To restore configurations

1Access the logic analyzers file system as the control user.

Refer to the chapter "Accessing the Logic Analyzer File System".

2Copy the setup.raw file to the appropriate directory.

For system configurations, this would be the \system directory. For
analyzer configurations, this would be the \slot_a directory. For
oscilloscope configurations, it would be the \slot_b directory. If you
copy a configuration file to a different directory than it came from, the
analyzer will not know how to interpret it.
Example Restoring an analyzer configuration
Suppose you want to load a configuration file called "486_bus" from
your local computer into the logic analyzers analyzer. The analyzer is
always in slot A. The logic analyzer is mounted on your network as disk
drive L:.
To load the configuration file, at the MS-DOS prompt enter:
copy 486_bus L:\slot_a\setup.raw
If your computer is running the UNIX operating system, you might use
the cp command. In an MSWindows environment, you can use File
Manager.