Logic Analyzer Reference
The Chart Menu
Min and Max scaling fields
When State is selected for the X axis, the minimum and maximum
values can range from -1 M to +1 M, depending on the trace point
location.
When Label is selected for either axis, the minimum and maximum
values range from 00000000 hex to FFFFFFFF hex regardless of the
axis, because labels are restricted to 32 bits.
Markers/Range field
The Marker/Range field is a toggle field. If the field is set to Range, X
and Y range fields become available to set the chart minimum and
maximum range points. If the field is set to Markers, a marker selection
menu appears with marker choices available with the present analyzer
configuration.
In a state analyzer with Count Off in the Trigger menu, only Pattern
markers are available. With Count Time on, Time markers and Statistics
markers become available. With Count States on, States markers are
available. The markers function just like those of the Waveform and
Listing displays.