511
Retrieving and Restoring Data Using the LAN
Retrieving and Restoring Data Using the LAN
To copy ASCII measurement data
1Set up the measurement you want to make, and run the analyzer
to acquire data.
For more information on setting up measurements, see the
Logic Analyzer section of this book.
2Access the logic analyzers file system.
Refer to the chapter "Accessing the Logic Analyzer File System".
You can do this from the XWindow interface, your computer using NFS
or ftp, from the front panel, or by programming the logic analyzer.
3Copy the measurement data from the \slot_a\data.asc
subdirectory.
In the analyzers file system, ASCII data files are located in
\slot_a\data.asc\{analyzer name}\{label name}.txt.
There is an ASCII data file corresponding to each label name you have
created in the measurement module. The data files for each label are
overwritten by the analyzer each time new data is acquired.
NOTE: 1670G-Series with the oscilloscope option only
Oscilloscope data is under \slot_b\data.asc. There is n o subdirectory matching
{analyzer name} but there are files for both oscilloscope channels.