150
Using the Oscilloscope
Calibrating the oscilloscope
2Optimize Delay of the Self Cal.
aObtain a BNC 50-W, 48-inch cable. Once you select Start, the
instrument will prompt you to connect the cable to the
appropriate location on the rear panel of the instrument.
bSelect the Procedure field, then select Delay from the pop-up
menu.
cSelect the Channel field, then select C1.
dSelect the Start field and follow the instructions on the
display.
eRepeat steps c and d for channel 2.
fAfter completing all of the channel delay calibrations, remove
the cable from the oscilloscope.
3Optimize the Time Null of the Self Cal.
aConnect two BNC 50-W, 9-inch cables to the BNC tee adapter.
Connect the BNC 50W (f)(f) adapter to the BNC tee adapter,
and connect the 48-inch BNC cable to the BNC 50W (f)(f)
adapter. Once you select Start, the instrument will prompt
you to connect the cables to the appropriate locations on the
rear panel of the instrument.
bSelect the Procedure field, then select Time N ull from the
pop-up menu.
cSelect the Start field and follow the instructions on the
display.
dAfter completion of the Time Null calibration, remove the
cables from the instrument.