261
Logic Analyzer Reference
Probing
Pattern generator output pod characteristics (pattern generator option only)
The following equivalent circuit information is provided to help you
select the appropriate clock and data pods for your application.

Agilent Technologies 10461A TTL Data Pod

Output type 10H125 with 100 ohm in series
Maximum clock 200 MHz
Skew Typical <2 ns; worst case 4 ns (note 1)
Recommended lead set Agilent Technologies 10474A

Agilent Technologies 10462A 3-State TTL/CMOS Data Pod

Output type 74ACT11244 with 100 ohm in series
10H125 on non 3-state channel 7 (see note 2)
3-state enable negative true, 100K ohm to GND
enabled on no connect
Maximum clock 100 MHz
Skew Typical <4 ns; worst case 12 ns (note 1)
Recommended lead set Agilent Technologies 10474A