441
Specifications
General Information
Specifications (oscilloscope option)
The specifications are the performance standards against which the
1670G-series logic analyzers oscilloscope is tested.
Bandwidth*: dc to 500 MHz (realtime, dc coupled)
Time Interval Measurement +/-[(0.005% of t) +
Accuracy*, **: (2 x 10-6 * delay setting) + 150 ps]
DC Offset Accuracy*: +/-(1.0% of channel offset + 2.0% of
full scale)
Voltage Measurement +/-((1.5% of full scale +
Accuracy*: offset accuracy + 0.016 V/div)
+ 0.008 * V/div )
Trigger Sensitivity*:
dc to 50 MHz: 0.063 * full scale
50 MHz to 500 MHz: 0.125 * full scale
Input R: 1 M +/-1%,50 +/-1%
* = Specifications valid within +/-10 °C of self-calibration temperature
** Specification applies at the maximum sample rate. At lower rates,
specification should be +/-(0.005% x t) + (2 x 10-6 x delay setting) +
(0.15 X sample interval) for bandwidth limited signals (tr = 1.4 x
sample interval). Sample interval is defined as .
1
samplerate
-----------------------------