332
Logic Analyzer Reference
The Listing Menu
The Listing Menu

Markers

The Markers field accesses the markers selection menu. When the
Markers field is selected, a marker selection menu appears with the
marker choices appropriate for the present analyzer configuration.
Off
The Off selection turns off marker operations but does not turn off
operations based on the markers. For example, if a stop measurement
was previously specified and the stop measurement criteria are met,
the measurement will stop even though the markers are off.

Pattern markers

Pattern markers identify and mark unique bit patterns in the data
listing. Once the unique bit patterns are marked, you can use them as
reference points or as criteria for a stop measurement.
When a marker is positioned in the Listing menu, it is also positioned in
the Chart menu and Waveform menu, but not in the Mixed Display
menu.

State analyzer markers

In a state analyzer with Count Off in the Trigger menu, only Pattern
markers are available. With Count Time turned on, Time markers and
Statistics markers become available. With Count States turned on,
State markers become available.