Using the Oscilloscope
The Scope Marker Menu
Slope fieldThe Slope field sets the Tx or To marker on either the positive or 
negative edge of the selected occurrence of a waveform. When you 
select the Slope field, the slope toggles between Positive and Negative. 
The default selection for the Slope field is Positive.
Occur fieldThe Occur field sets the Tx or To marker on a specific occurrence of a 
displayed edge on the waveform. You can define the edge to be 
displayed all the way from the 1st edge up to the 100th edge. The 
count of edge occurrences is made starting with the first edge 
displayed on the screen. 
Auto-marker measurements are made with data that is displayed on 
the screen. Make sure the data of interest is fully displayed on the 
screen.
Any number from 1 to 100 in increments of 1 can be entered. The 
default value for the Occur field is 1. 
Statistics fieldThe Statistics field allows you to make minimum, maximum, and mean 
time interval measurements from marker Tx to marker To. When you 
select the Statistics field, it toggles between On and Off. The default 
selection for the Statistics field is On.
OnWhen Statistics is set to On, the minimum, maximum, and mean 
(average) Tx to To marker time interval data is accumulated and 
displayed to the right of the T Markers field until one of the following 
happens:
•Auto is deselected as the time marker option.
•Auto-marker parameters are changed.
•Statistics is set to Off. 
•Run Repetitive mode is stopped.