Using the Oscilloscope
The Scope Marker Menu
Slope field
The Slope field sets the Tx or To marker on either the positive or
negative edge of the selected occurrence of a waveform. When you
select the Slope field, the slope toggles between Positive and Negative.
The default selection for the Slope field is Positive.
Occur field
The Occur field sets the Tx or To marker on a specific occurrence of a
displayed edge on the waveform. You can define the edge to be
displayed all the way from the 1st edge up to the 100th edge. The
count of edge occurrences is made starting with the first edge
displayed on the screen.
Auto-marker measurements are made with data that is displayed on
the screen. Make sure the data of interest is fully displayed on the
screen.
Any number from 1 to 100 in increments of 1 can be entered. The
default value for the Occur field is 1.
Statistics field
The Statistics field allows you to make minimum, maximum, and mean
time interval measurements from marker Tx to marker To. When you
select the Statistics field, it toggles between On and Off. The default
selection for the Statistics field is On.
On
When Statistics is set to On, the minimum, maximum, and mean
(average) Tx to To marker time interval data is accumulated and
displayed to the right of the T Markers field until one of the following
happens:
Auto is deselected as the time marker option.
Auto-marker parameters are changed.
Statistics is set to Off.
Run Repetitive mode is stopped.