Troubleshooting the LAN Connection
Solutions to Common Problems
If you cannot copy files from the logic analyzer
If you can only copy a few bytes of a file:
Copy the file of interest to your PC or workstation, and use the new, local
copy as your working copy. Some applications cannot work directly with
the dynamic files from the logic analyzer.
If you are on a UNIX workstation, try using dd instead of cp.
If you cannot restore raw files
If the setup.raw and data.raw files seem to be ignored when you copy
them to the logic analyzer:
Verify the LAN connection between your computer and the log ic analyzer.
Refer to "If you cannot connect to the logic analyzer" earlier in this section.
Change the file time-out. To change the time-out:

1Go to the logic analyzer System External I/O menu.

2Select the LAN Settings field.

3Set the file time out.

A value of about 1.5 seconds should be sufficient. Set the time out to
the minimum value that will allow network operations to work reliably.