325
Logic Analyzer Reference
The Analyzer Trigger Menu
Using Preset Values
Assign. Assign toggles which machine the term is assigned to. All of
the available resource terms except the Edge terms can be assigned to
any analyzer. However, a term can only be assigned to one analyzer at a
time.
Rename. Rename lets you change the term name. This function works
for all terms.
Clear (=X). Clear sets the terms to their broadest possible meaning.
For terms a - j, the assignment field is set to all Xs (dont cares). For
Range 1 and 2, the boundaries are set to the maximum and minimum
values. For Timers 1 and 2, the assignment field is set to a minimum
time of 400 ns. For Edge 1 and 2, the assignment field is set to dont
cares.
Set (=1). In terms a - j, the assignment field is set to its maximum
value, with all bits set to 1. This option is not available for the Range,
Timer, and Edge terms.
Reset (=0). In terms a - j, the assignment field is set to its minimum
value, with all bits set to 0. This option is not available for the Range,
Timer, and Edge terms.
Assigning Bit by Bit
Bit pattern terms. Just to the right of the bit pattern name fields are
the term assignment fields. When any of the individual assignment
fields are selected, a keypad appears. Use this keypad to assign real
values or Dont Care (X) values.
Edge terms. Assign edge terms the same way you do bit pattern
terms. Edge terms can be used singularly or in combination with each
other across all assigned channels. When you specify an edge on more
than one channel, the analyzer ORs the edges.