Logic Analyzer Reference
The Mixed Display Menu
The Mixed Display Menu
The Mixed Display menu combines a state listing display located at the
top of the menu and a waveform display located at the bottom of the
menu. The Mixed Display menu shows both state and timing data in the
same display.
The Mixed Display menu only becomes available when at least one
analyzer is configured as a state analyzer, with its Count field in the
Trigger menu set to Time. If two state analyzers are configured, both
state listing displays can be interleaved as well as shown separately,
however, the listing menus are the best display menus for a two-state
analyzer configuration.
The waveform display area shows timing analyzer waveforms from the
configured timing analyzer, the oscilloscope waveform
(analyzers with the oscilloscope option), or both.

Interleaving state listings

Interleaved state listings lets you view two labels and their data from
different analyzers in the same column. The process of interleaving
state listings can be performed in either the Listing menu or the Mixed
Display menu. For example, if data is interleaved in the Listing menu, it
will be automatically interleaved in the Mixed Display menu.
The interleaved label is placed directly above the selected label, and all
interleaved data is displayed in yellow. In addition, the state numbers
of the interleaved data are indented to the right. Because of the lack of
room available in the listing portion of the Mixed Display menu, the
label identifying the interleaved data is not displayed. For this reason,
when two state analyzers are configured the listing menus should be
used to display interleaved labels.