305
Logic Analyzer Reference
The Analyzer Format Menu
Master and Slave Clock fields (State only)
The Master and Slave Clock fields are used to construct a clocking
arrangement. A clocking arrangement is the assignment of appropriate
clocks, clock edges, and clock qualifier levels which allow the analyzer
to synchronize itself on valid data.

Clock selections

When the Master or Slave Clock field is selected, a clock/qualifier
selection menu appears showing the available clocks and qualifiers for
a clocking arrangement. Depending on the model, there are a
maximum of four clocks available (J through M), and a maximum of
four clock qualifiers available (Q1 through Q4).
Clock edges are ORed to clock edges, clock qualifiers are ANDed to
clock edges, and clock qualifiers can be either ANDed or ORed
together. All clock and qualifier combinations on the left side of the
graphic line are ORed to all combinations on the right side of the line.
For example, in a six-clock model, all combinations of the J, K, and L
clock with Q1 and Q2 qualifiers, are ORed to the clock combinations of
the M, N, and P clocks with Q3 and Q4 qualifiers.