335
Logic Analyzer Reference
The Waveform Menu
Delay field
Depending on the analyzer configuration, a positive or negative delay
measured in either states (State only) or time (Timing only) can be set.
The Delay field lets you scroll the data and place the display window at
center screen. Changing the delay will not affect the data acquisition
unless it is a timing analyzer and the acquisition mode is automatic. In
this case, the sample period changes.
The delay range of a timing analyzer is from -2500 seconds to +2500
seconds. The delay range of a state analyzer is dependent on memory
length and cannot exceed total memory size.
Waveform label field
The waveform label field, located on the left side of the waveform
display, is both a display and configuration field. After all desired
waveforms are configured for display, they are listed in the waveform
label field. If there are more waveforms than can be displayed, you can
roll the list by selecting the waveform label field, then after the roll
indicator appears, turn the knob.
If the waveform label field is selected a second time, a waveform
modification menu appears. Use this menu to configure the waveform
display.
When the waveform modification menu appears, select the operation
to insert, replace, delete, or scale waveforms into the display. You can
display a maximum of 24 waveforms on screen at one time.