368
System Performance Analysis (SPA) Software
System Performance Analysis Software
Min, Max, and Avg Time Statistics. The Time Interval mode display
shows three statistics: Maximum (Max) time, Minimum (Min) time,
and Average (Avg) time. These values are displayed whether or not
they fall into any of the time interval ranges. Therefore, they are
helpful in determining if the appropriate time intervals have been
chosen.
The maximum resolution of the statistics is 8 ns.
Total samples. Total samples is displayed above the histogram. A
sample is defined as one Start/End pair. The Total samples field is not
updated until the analyzers memory is full or until Stop is pressed.