Agilent Technologies E4406A VSA Example, CALCulate:SPECtrum:MARKer2:MAXimum, Calc:Spec:Mark:Aoff

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 204
Image 204
Example:

Language Reference

CALCulate Subsystem

CHPower - no markers

EVMQpsk - markers available

PSTatistic - markers available

RHO - markers available

SPECtrum - markers available

WAVeform - markers available

Example:

Suppose you are using the Spectrum measurement. To position marker 2 at the maximum peak value of the trace that marker 2 is currently on, the command is:

:CALCulate:SPECtrum:MARKer2:MAXimum

You must make sure that the measurement is completed before trying to query the marker value. Use the MEASure or READ command before using the marker command. This forces the measurement to complete before allowing the next command to be executed.

Markers All Off on All Traces

:CALCulate:<measurement>:MARKer:AOFF

Turns off all markers on all the traces in the specified measurement.

Example:

CALC:SPEC:MARK:AOFF

Remarks:

The keyword for the current measurement must be

 

specified in the command. (Some examples include:

 

SPECtrum, WAVeform)

Front Panel

 

Access:

Marker, More, Marker All Off

Marker Function

:CALCulate:<measurement>:MARKer[1]234:FUNCtion

BPOWerNOISeOFF

:CALCulate:<measurement>:MARKer[1]234:FUNCtion?

Selects the type of marker for the specified marker. A particular measurement may not have all the types of markers that are commonly available.

The marker must have already been assigned to a trace. Use :CALCulate:<measurement>:MARKer[1]234:TRACe to assign a marker to a particular trace.

204

Chapter 5

Page 204
Image 204
Agilent Technologies E4406A VSA manual Example, CALCulate:SPECtrum:MARKer2:MAXimum, Markers All Off on All Traces