Agilent Technologies E4406A VSA manual Questionable Integrity Register

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 372
Image 372
Questionable Frequency Negative Transition

Language Reference

STATus Subsystem

Questionable Frequency Negative Transition

:STATus:QUEStionable:FREQuency:NTRansition <number>

:STATus:QUEStionable:FREQuency:NTRansition?

This command determines what bits in the Questionable Frequency Condition register will set the corresponding bit in the Questionable Frequency Event register when the condition register bit has a negative transition (1 to 0). The variable <number> is the sum of the decimal values of the bits that you want to enable.

Key Type:

There is no equivalent front panel key.

Factory Preset

 

and *RST:

0

Range:

0 to 32767

Questionable Frequency Positive Transition

:STATus:QUEStionable:FREQuency:PTRansition <number>

:STATus:QUEStionable:FREQuency:PTRansition?

This command determines what bits in the Questionable Frequency Condition register will set the corresponding bit in the Questionable Frequency Event register when the condition register bit has a positive transition (0 to 1). The variable <number> is the sum of the decimal values of the bits that you want to enable.

Key Type:

There is no equivalent front panel key.

Factory Preset

 

and *RST:

32767 (all 1’s)

Range:

0 to 32767

Questionable Integrity Register

Questionable Integrity Condition :STATus:QUEStionable:INTegrity:CONDition?

This query returns the decimal value of the sum of the bits in the

Questionable Integrity Condition register.

NOTE

The data in this register is continuously updated and reflects the

 

current conditions.

 

Key Type:

There is no equivalent front panel key.

372

Chapter 5

Page 372
Image 372