Power vs. Time—Averaging Mode

Language Reference

SENSe Subsystem

Power vs. Time—Averaging Mode

[:SENSe]:PVTime:AVERage:TCONtrol EXPonentialREPeat

[:SENSe]:PVTime:AVERage:TCONtrol?

Select the type of termination control used for the averaging function. This specifies the averaging action after the specified number of bursts (average count) is reached.

EXPonential - Each successive data acquisition after the average count is reached is exponentially weighted and combined with the existing average.

REPeat - After reaching the average count, the averaging is reset and a new average is started.

Factory Preset

 

and *RST:

EXPonential

Remarks:

You must be in the EDGE(w/GSM), GSM or Service

 

mode to use this command. Use INSTrument:SELect to

 

set the mode.

Power vs. Time—Averaging Type

[:SENSe]:PVTime:AVERage:TYPE

LOGMAXimumMINimumMXMinimumRMS

[:SENSe]:PVTime:AVERage:TYPE?

Select the type of averaging to be performed.

LOG - The log of the power is averaged. (This is also known as video averaging.)

MAXimum - The maximum values are retained.

MINimum - The minimum values are retained.

MXMinimum - Both the maximum and the minimum values are retained.

RMS - The power is averaged, providing the rms of the voltage.

Factory Preset

 

and *RST:

RMS

Remarks:

You must be in the EDGE(w/GSM), GSM or Service

 

mode to use this command. Use INSTrument:SELect to

 

set the mode.

336

Chapter 5

Page 336
Image 336