Align the Trigger Delay

Language Reference

CALibration Subsystem

Align the Trigger Delay

:CALibration:TRIGger:DELay

:CALibration:TRIGger:DELay?

Align any trigger delays needed. One place that this alignment is used is for the even second clock functionality in cdmaOne mode. This same alignment is run as part of the CAL:ALL routine.

The query performs the alignment and returns a zero if the alignment is successful.

Front Panel

Access: System, Alignments, Align subsystem, Align 50 MHz

Reference

Align the Trigger Interpolator

:CALibration:TRIGger:INTerpolator

:CALibration:TRIGger:INTerpolator?

Align the partial sample trigger interpolator. This same alignment is run as part of the CAL:ALL routine.

The query performs the alignment and returns a zero if the alignment is successful.

Front Panel

Access: System, Alignments, Align subsystem, Align 50 MHz

Reference

Calibration Wait

:CALibration:WAIT

Waits until any alignment procedure that is underway is completed.