Agilent Technologies E4406A VSA SENSe:CHPower:SWEep:TIME:AUTO OFF|ON|0|1, Channel Power—SweepTime

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 328
Image 328
[:SENSe]:CHPower:SWEep:TIME:AUTO OFFON01

Language Reference

SENSe Subsystem

Range:

1 ∝s to 50 ms

Default Unit:

seconds

Remarks:

You must be in Basic, cdmaOne, cdma2000, W-CDMA

 

(3GPP), or W-CDMA (Trial & Arib) mode to use this

 

command. Use INSTrument:SELect to set the mode.

History:

Version A.03.00 and later

Channel Power—Sweep Time

[:SENSe]:CHPower:SWEep:TIME:AUTO OFFON01

[:SENSe]:CHPower:SWEep:TIME:AUTO?

Selects the automatic sweep time, optimizing the measurement.

Factory Preset

 

and *RST:

ON

Remarks:

You must be in Basic, cdmaOne, cdma2000, W-CDMA

 

(3GPP), or W-CDMA (Trial & Arib) mode to use this

 

command. Use INSTrument:SELect to set the mode.

History:

Version A.03.00 and later

Channel Power—Trigger Source

[:SENSe]:CHPower:TRIGger:SOURce

EXTernal[1]EXTernal2IMMediate

[:SENSe]:CHPower:TRIGger:SOURce?

Select the trigger source used to control the data acquisitions. This is an Advanced control that normally does not need to be changed.

EXTernal 1 - front panel external trigger input

EXTernal 2 - rear panel external trigger input

IMMediate - the next data acquisition is immediately taken (also called Free Run).

Factory Preset

 

and *RST:

IMMediate

Remarks:

You must be in the cdmaOne, cdma2000, W-CDMA

 

(3GPP), W-CDMA (Trial & Arib), or Basic mode to use

 

this command. Use INSTrument:SELect to set the

 

mode.