Agilent Technologies E4406A VSA manual Adjacent Channel Power—PowerReference, Baseband IQ Commands

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 317
Image 317
Adjacent Channel Power—Power Reference

 

Language Reference

 

SENSe Subsystem

 

RFBurst for MS

Remarks:

You must be in Basic, cdmaOne, iDEN, NADC, or PDC

 

mode to use this command. Use INSTrument:SELect to

 

set the mode.

 

In Basic mode, for offset frequencies >12.5 MHz, the

 

external triggers will be a more reliable trigger source

 

than RF burst. Also, you can use the Waveform

 

measurement to set up trigger delay.

Adjacent Channel Power—Power Reference

[:SENSe]:ACP:TYPE PSDRefTPRef

[:SENSe]:ACP:TYPE?

Selects the measurement type. This allows you to make absolute and relative power measurements of either total power or the power normalized to the measurement bandwidth.

Power Spectral Density Reference (PSDRef) - the power spectral density is used as the power reference

Total Power Reference (TPRef) - the total power is used as the power reference

Factory Preset

 

and *RST:

Total power reference (TPRef)

Remarks:

You must be in the Basic, cdmaOne, cdma2000,

 

W-CDMA (3GPP), W-CDMA (Trial & Arib), NADC, or

 

PDC mode to use this command. Use

 

INSTrument:SELect to set the mode.

Baseband IQ Commands

Baseband I/Q - Select I/Q Power Range

[:SENSe]:POWer:IQ:RANGe[:UPPer] <power> [DBM]DBMVW

[:SENSe]:POWer:IQ:RANGe[:UPPer]?

Selects maximum total power expected from unit under test at test port when I or Q port is selected.

Range:

For 50 Ohms:

13.0, 7.0. 1.0, or 5.1 dBm 60.0, 54.0, 48.0, or 41.9 dBmV 0.02, 0.005, 0.0013, or 0.00031 W