Agilent Technologies E4406A VSA SENSe:ACP:FILTer:RRC:ALPHa <numeric>, SENSe:ACP:FILTer:RRC:ALPHa?

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 291
Image 291
[:SENSe]:ACP:FILTer[:RRC]:ALPHa <numeric>

Language Reference

SENSe Subsystem

measure the offset channels and wider dynamic range for the measurement is available.

Factory Preset

 

and *RST:

0

Range:

40.00 to 0.00 dB

Remarks:

You must be in the W-CDMA (3GPP) mode to use this

 

command. Use INSTrument:SELect to set the mode.

Adjacent Channel Power—Root Raised Cosine Filter Alpha

[:SENSe]:ACP:FILTer[:RRC]:ALPHa <numeric>

[:SENSe]:ACP:FILTer[:RRC]:ALPHa?

Set the alpha value of the Root Raised Cosine (RRC) filter.

Factory Preset

 

and *RST:

0.22

Range:

0.01 to 0.5

Remarks:

You must be in the W-CDMA (3GPP) mode to use this

 

command. Use INSTrument:SELect to set the mode.

Adjacent Channel Power—Root Raised Cosine Filter Control

[:SENSe]:ACP:FILTer[:RRC][:STATe] OFFON01

[:SENSe]:ACP:FILTer[:RRC][:STATe]?

Turn the Root Raised Cosine (RRC) filter on or off.

Factory Preset

 

and *RST:

On

Remarks:

You must be in the W-CDMA (3GPP) mode to use this

 

command. Use INSTrument:SELect to set the mode.

Adjacent Channel Power—Reference Channel FFT Segments

[:SENSe]:ACP:FFTSegment <integer>

[:SENSe]:ACP:FFTSegment?

Selects the number of FFT segments used in making the measurement of the reference channel (carrier). In automatic mode the measurement optimizes the number of FFT segments required for the shortest measurement time. The minimum number of segments required to make a measurement is set by your desired measurement bandwidth.

Chapter 5

291

Page 291
Image 291