Agilent Technologies E4406A VSA manual Waveform—AveragingMode, SENSe:WAVeform:AVERage:TCONtrol?

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 360
Image 360
Waveform—Averaging Mode

Language Reference

SENSe Subsystem

selected with INSTrument:SELect.

Waveform—Averaging Mode

[:SENSe]:WAVeform:AVERage:TCONtrol EXPonentialREPeat

[:SENSe]:WAVeform:AVERage:TCONtrol?

Select the type of termination control used for the averaging function. This determines the averaging action after the specified number of ‘sweeps’ (average count) is reached.

EXPonential - Each successive data acquisition after the average count is reached, is exponentially weighted and combined with the existing average.

REPeat - After reaching the average count, the averaging is reset and a new average is started.

Factory Preset

 

and *RST:

EXPonential

Remarks:

To use this command, the appropriate mode should be

 

selected with INSTrument:SELect.

Waveform—Averaging Type

[:SENSe]:WAVeform:AVERage:TYPE

LOGMAXimumMINimumRMSSCALar

[:SENSe]:WAVeform:AVERage:TYPE?

Select the type of averaging.

LOG - The log of the power is averaged. (This is also known as video averaging.)

MAXimum - The maximum values are retained.

MINimum - The minimum values are retained.

RMS - The power is averaged, providing the rms of the voltage.

Factory Preset

 

and *RST:

RMS

Remarks:

To use this command, the appropriate mode should be

 

selected with INSTrument:SELect.

Waveform—Resolution BW