[:SENSe]:ACP:LIST:ALIMit

Language Reference

SENSe Subsystem

Adjacent Channel Power—Offset Frequency Absolute Limit

[:SENSe]:ACP:LIST:ALIMit

<abs_powr>,<abs_powr>,<abs_powr>,<abs_powr>,<abs_powr>

[:SENSe]:ACP:LIST:ALIMit?

Set the absolute limit on offset frequencies relative to the carrier. You can turn off (not use) specific offsets with the [:SENSe]:ACP:LIST:STATe command.

Factory Preset and *RST:

Mode

Offset A

Offset B

Offset C

Offset D

Offset E

 

 

 

 

 

 

NADC

0 dBm

0 dBm

13 dBm

0 dBm

0 dBm

 

 

 

 

 

 

PDC

0 dBm

0 dBm

0 dBm

0 dBm

0 dBm

 

 

 

 

 

 

Range:

200 to 50 dBm

Remarks:

You must be in the NADC, cdmaOne, or PDC mode to

 

use this command. Use INSTrument:SELect to set the

 

mode.

Adjacent Channel Power—Offset Frequency

[:SENSe]:ACP:LIST[:FREQuency]

<f_offset>,<f_offset>,<f_offset>,<f_offset>,<f_offset>

[:SENSe]:ACP:LIST[:FREQuency]?

Define the offset frequencies. You can turn off (not use) specific offsets with the [:SENSe]:ACP:LIST:STATe command.

Factory Preset and *RST:

Mode

Offset A

Offset B

Offset C

Offset D

Offset E

 

 

 

 

 

 

NADC

30 kHz

60 kHz

90 kHz

120 kHz

0 Hz

 

 

 

 

 

 

PDC

50 kHz

100 kHz

0 kHz

0 kHz

0 kHz

 

 

 

 

 

 

Range:

10 Hz to 45 MHz

 

 

 

 

0 to 200 kHz

 

 

 