Variable Parameters

Programming Fundamentals

SCPI Language Basics

Variable Parameters

<freq>

 

<bandwidth>

Is a positive rational number followed by optional units.

 

The default unit is Hz. Acceptable units include: HZ,

 

KHZ, MHZ, GHZ.

<time>

 

<seconds>

Is a rational number followed by optional units. The

 

default units are seconds. Acceptable units include: S,

 

MS, US.

<voltage>

Is a rational number followed by optional units. The

 

default units are V. Acceptable units include: Volts, V,

 

MV, UV.

<power>

 

<ampl>

Is a rational number followed by optional units. The

 

default units are dBm. Acceptable units include: DBM,

 

DBMV, W.

<rel_power>

 

<rel_ampl>

Is a positive rational number followed by optional units.

 

The default units are dB. Acceptable units include: DB.

<angle>

 

<degrees>

Is a rational number followed by optional units. The

 

default units are degrees. Acceptable units include:

 

DEG, RAD.

<integer>

An integer value has no units.

<percent>

Is a rational number between 0 and 100, with no units.

<string>

Is a series of alpha numeric characters.

<bit_pattern>

Specifies a series of bits rather than a numeric value.

 

The bit series is the binary representation of a numeric

 

value. There are no units.

 

Bit patterns are most often specified as hexadecimal

 

numbers, though octal, binary or decimal numbers may

 

also be used. In the SCPI language these numbers are

 

specified as:

 

Hexadecimal, #Hdddd or #hdddd where ‘d’ represents a

 

hexadecimal digit 0 to 9 and ‘a’ to ‘f ’. So #h14 can be used

 

instead of the decimal number 20.

 

• Octal, #Odddddd or #odddddd where ‘d’ represents an

 

octal digit 0 to 7. So #o24 can be used instead of the

 

decimal number 20.

 

Binary, #Bdddddddddddddddd or #bdddddddddddddddd

 

where ‘d’ represents a 1 or 0. So #b10100 can be used

 

instead of the decimal number 20.

Chapter 2

63