Align the Wide LC Prefilter

 

Language Reference

 

CALibration Subsystem

Remarks:

A valid service password needs to be entered prior to

 

sending the command.

Front Panel

 

Access:

System, Alignments, Align Subsystem, IF

Align the Wide LC Prefilter

:CALibration:PFILter:LC:WIDE

:CALibration:PFILter:LC:WIDE?

Align the wide LC prefilter. (1.2 MHz to 7.5 MHz)

The query performs the alignment and returns a zero if the alignment is successful.

Remarks:

A valid service password needs to be entered prior to

 

sending the command.

Front Panel

 

Access:

System, Diagnostics

Align the Narrow Crystal Prefilter

:CALibration:PFILter:XTAL:NARRow

:CALibration:PFILter:XTAL:NARRow?

Align the narrow crystal prefilter. (2.5 kHz to 20 kHz)

The query performs the alignment and returns a zero if the alignment is successful.

Remarks:

A valid service password needs to be entered prior to

 

sending the command.

Front Panel

 

Access:

Enter service password and press System, Diagnostics

Align the Wide Crystal Prefilter

:CALibration:PFILter:XTAL:WIDE

:CALibration:PFILter:XTAL:WIDE?

Align the wide crystal prefilter. (20 kHz to 200 kHz)

The query performs the alignment and returns a zero if the alignment is successful.

Chapter 5

221