Agilent Technologies E4406A VSA manual CALibration:REF50:ENTer, CALibration:REF50:EXIT

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 224
Image 224
:CALibration:REF50:ENTer

Language Reference

CALibration Subsystem

A valid service password needs to be entered prior to sending this command.

Front Panel

Access: System, Alignments, Align subsystem, Align 50 MHz

Reference

Enter Interactive Mode for Internal 50 MHz Amplitude Reference Alignment

:CALibration:REF50:ENTer

Turns on the interactive mode for alignment of the internal 50 MHz amplitude reference signal. Use CAL:REF50:ANOW to do the alignment and CAL:REF50:EXIT to exit the interactive mode.

Remarks:

You must be in the Service mode to use this command.

 

Use INSTrument:SELect.

 

A valid service password needs to be entered prior to

 

sending this command.

Front Panel

System, Alignments, Align subsystem, Align 50 MHz

Access:

 

Reference

Exit Interactive Mode for Internal 50 MHz Amplitude Reference Alignment

:CALibration:REF50:EXIT

Turns off the interactive mode for alignment of the internal 50 MHz amplitude reference signal. Use CAL:REF50:ENTer to turn the mode on and CAL:REF50:ANOW to do the alignment immediately.

Remarks:

You must be in the Service mode to use this command.

 

Use INSTrument:SELect.

 

A valid service password needs to be entered prior to

 

sending the command.

Front Panel

System, Alignments, Align subsystem, Align 50 MHz

Access:

 

Reference

Query the Absolute Level for the 50 MHz Amplitude Reference

:CALibration:REF50:LAST:ALEVel?

Query returns the last value of the absolute level of the 50 MHz reference alignment.

224

Chapter 5

Page 224
Image 224