Programming Fundamentals

Using the Instrument Status Registers

The standard event status register contains the following bits:

Bit

Description

 

 

0

A 1 in this bit position indicates that all pending operations were completed following

 

execution of the *OPC command.

 

 

1

This bit is always set to 0. (The instrument does not request control.)

 

 

2

A 1 in this bit position indicates that a query error has occurred. Query errors have SCPI

 

error numbers from 499 to 400.

 

 

3

A 1 in this bit position indicates that a device dependent error has occurred. Device

 

dependent errors have SCPI error numbers from 399 to 300 and 1 to 32767.

 

 

4

A 1 in this bit position indicates that an execution error has occurred. Execution errors

 

have SCPI error numbers from 299 to 200.

 

 

5

A 1 in this bit position indicates that a command error has occurred. Command errors

 

have SCPI error numbers from 199 to 100.

 

 

6

Currently not used.

 

 

7

A 1 in this bit position indicates that the instrument has been turned off and then on.

 

 

The standard event status register is used to determine the specific event that set bit 5 in the status byte register. To query the standard event status register, send the command *ESR?. The response will be the decimal sum of the bits which are enabled (set to 1). For example, if bit number 7 and bit number 3 are enabled, the decimal sum of the 2 bits is 128 plus 8. So the decimal value 136 is returned.

Chapter 2

87

Page 87