:CALibration:REF50:ANOW

Language Reference

CALibration Subsystem

This is used for aligning the 50 MHz amplitude reference with CAL:REF50.

Preset

25.00 dBm

and *RST:

Range:

30 to 20 dBm

Default Unit:

dBm

Remarks:

You must be in the Service mode to use this command.

 

Use INSTrument:SELect.

 

A valid service password needs to be entered prior to

 

sending this command.

Front Panel

System, Alignments, Align subsystem, Align 50 MHz

Access:

 

Reference

Internal 50 MHz Amplitude Reference Alignment Control

:CALibration:REF50:ANOW

Immediately does the automatic alignment of the internal 50 MHz amplitude reference oscillator. This command is used with the interactive mode of the 50 MHz alignment, i.e. CAL:REF50:ENTer.

Remarks:

You must be in the Service mode to use this command.

 

Use INSTrument:SELect.

 

A valid service password needs to be entered prior to

 

sending this command.

Front Panel

System, Alignments, Align subsystem, Align 50 MHz

Access:

 

Reference

Internal 50 MHz Amplitude Reference Alignment Control

:CALibration:REF50[:DOIT]

:CALibration:REF50[:DOIT]?

Does automatic alignment of the internal 50 MHz amplitude reference oscillator. You do this by setting an external source to 25.00 dBm and using a power meter to measure the exact value. Then use CAL:REF50:AMPL to input the source amplitude, measured on the power meter. Finally, connect the source to the instrument RF INPUT port and run the adjustment.

Remarks:

You must be in the Service mode to use this command.

 

Use INSTrument:SELect.