Agilent Technologies E4406A VSA manual Select Display Format, Spectrum - Y-AxisScale/Div

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 230
Image 230
Select Display Format

Language Reference

DISPlay Subsystem

1.If in local, press any key

2.If in remote, press the local (system) key

3.If in local lockout, no key

Front Panel

Access: System, Disp Updates for VSA

Select Display Format

:DISPlay:FORMat:TILE

Selects the viewing format that displays multiple windows of the current measurement data simultaneously. Use DISP:FORM:ZOOM to return the display to a single window.

Front Panel

Access: Zoom (toggles between Tile and Zoom)

Select Display Format

:DISPlay:FORMat:ZOOM

Selects the viewing format that displays only one window of the current measurement data (the current active window). Use DISP:FORM:TILE to return the display to multiple windows.

Front Panel

Access: Zoom (toggles between Tile and Zoom)

Spectrum - Y-Axis Scale/Div

:DISPlay:SPECtrum[n]:WINDow[m]:TRACe:Y[:SCALe]:PDIVision

<power>

:DISPlay:SPECtrum[n]:WINDow[m]:TRACe:Y[:SCALe]:PDIVision?

Sets the amplitude reference level for the y-axis. n selects the view, the default is Spectrum.

n=1, m=1 Spectrum

n=1, m=2 I/Q Waveform

n=1, m=2 I and Q Waveform (Basic, W-CDMA, cdma2000)

n=1, m=3 numeric data (Service mode)

n=1, m=4 RF envelope (Service mode)