Agilent Technologies E4406A VSA manual Spectrum—Pre-FFTBW Filter Type, Spectrum—ResolutionBW

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 351
Image 351
Spectrum—Pre-FFT BW Filter Type

Language Reference

SENSe Subsystem

combinations of these settings that are not valid.

Factory Preset

 

and *RST:

1.55 MHz

 

1.25 MHz for cdmaOne

 

155.0 kHz, for iDEN mode

Range:

1 Hz to 10.0 MHz

Remarks:

To use this command, the appropriate mode should be

 

selected with INSTrument:SELect.

Spectrum—Pre-FFT BW Filter Type

[:SENSe]:SPECtrum:BANDwidthBWIDth:PFFT:TYPE FLATGAUSsian

[:SENSe]:SPECtrum:BANDwidthBWIDth:PFFT:TYPE?

Select the type of pre-FFT filter that is used. This is an advanced control that normally does not need to be changed.

Flat top (FLAT)- a filter with a flat amplitude response, which provides the best amplitude accuracy.

GAUSsian - a filter with Gaussian characteristics, which provides the best pulse response.

Factory Preset

 

and *RST:

FLAT

Remarks:

To use this command, the appropriate mode should be

 

selected with INSTrument:SELect.

Spectrum—Resolution BW

[:SENSe]:SPECtrum:BANDwidthBWIDth[:RESolution] <freq>

[:SENSe]:SPECtrum:BANDwidthBWIDth[:RESolution]?

Set the resolution bandwidth for the FFT. This is the bandwidth used for resolving the FFT measurement. It is not the pre-FFT bandwidth. This value is ignored if the function is auto-coupled.

Frequency span, resolution bandwidth, and the pre-FFT bandwidth settings are normally coupled. If you are not auto-coupled, there can be combinations of these settings that are not valid.

Factory Preset

 

and *RST:

20.0 kHz

 

250.0 Hz, for iDEN mode

Range:

0.10 Hz to 3.0 MHz