Agilent Technologies E4406A VSA manual Reference Oscillator Commands, SENSe:ROSCillator:OUTPut?

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 345
Image 345
Reference Oscillator Commands

 

Language Reference

 

SENSe Subsystem

Remarks:

You must be in the NADC or PDC mode to use this

 

command. Use INSTrument:SELect to set the mode.

Reference Oscillator Commands

Reference Oscillator External Frequency

[:SENSe]:ROSCillator:EXTernal:FREQuency <frequency>

[:SENSe]:ROSCillator:EXTernal:FREQuency?

Specify to the frequency of the external reference being supplied to the instrument. Switch to the external reference with ROSC:SOUR.

Preset

 

and *RST:

Value remains at last user selected value (persistent)

 

Factory default, 10 MHz

Range:

1 MHz to 30 MHz, with 1 Hz steps

Default Unit:

Hz

Remarks:

Global to system

Front Panel

 

Access:

System, Reference, Ref Oscillator

Reference Oscillator Rear Panel Output

[:SENSe]:ROSCillator:OUTPut[:STATe] OFFON01

[:SENSe]:ROSCillator:OUTPut?

Turn on and off the 10 MHz frequency reference signal going to the rear panel.

ESA? - Option oscillator commands, if applicable, are found as SENSe:OPTion:ROSCillator.

Preset

 

and *RST:

Persistent State with factory default of On

Remarks:

Global to system. Was SENS:ROSC:REAR

Front Panel

 

Access:

System, Reference, 10 MHz Out

Chapter 5