Power vs. Time—Resolution BW

Language Reference

SENSe Subsystem

Power vs. Time—Resolution BW

[:SENSe]:PVTime:BANDwidthBWIDth[:RESolution] <freq>

[:SENSe]:PVTime:BANDwidthBWIDth[:RESolution]?

Set the resolution BW. This is an advanced control that normally does not need to be changed. Setting this to a value other than the factory default, may cause invalid measurement results.

Factory Preset

 

and *RST:

500 kHz

Range:

1 kHz to 5 MHz

Default Unit:

Hz

Remarks:

You must be in the EDGE(w/GSM), GSM or Service

 

mode to use this command. Use INSTrument:SELect to

 

set the mode.

Power vs. Time—RBW Filter Type

[:SENSe]:PVTime:BANDwidthBWIDth[:RESolution]:TYPE

FLATtopGAUSsian

[:SENSe]:PVTime:BANDwidthBWIDth[:RESolution]:TYPE?

Select the type of resolution BW filter. This is an advanced control that normally does not need to be changed. Setting this to a value other than the factory default, may cause invalid measurement results.

FLATtop - a filter with a flat amplitude response, which provides the best amplitude accuracy.

GAUSsian - a filter with Gaussian characteristics, which provides the best pulse response.

Factory Preset

 

and *RST:

GAUSsian

Remarks:

You must be in the EDGE(w/GSM), GSM or Service

 

mode to use this command. Use INSTrument:SELect to

 

set the mode.

Power vs. Time—Sweep Time

[:SENSe]:PVTime:SWEep:TIME <integer>

[:SENSe]:PVTime:SWEep:TIME?

Set the number of slots which are used in each data acquisition. Each slot is approximately equal to 570 ms. The measurement is made for a small additional amount of time (about 130 ∝s) in order to view the