Image 214
Align the ADC Offset

Language Reference

CALibration Subsystem

Align the ADC Offset

:CALibration:ADC:OFFSet

:CALibration:ADC:OFFSet?

Align the six ADC offset DACs. This same alignment is run as part of the CAL:ALL routine.

The query performs the alignment and returns a zero if the alignment is successful.

Front Panel

Access: System, Alignments, Align subsystem, Align ADC

Align the ADC RAM Gain

:CALibration:ADCRam:GAIN

:CALibration:ADCRam:GAIN?

Align the gain of the six ADC RAM pages. This same alignment is run as part of the CAL:ALL routine.

The query performs the alignment and returns a zero if the alignment is successful.

Front Panel

Access: System, Alignments, Align subsystem, Align ADC???

Align All Instrument Assemblies

:CALibration[:ALL]

:CALibration[:ALL]?

Performs an alignment of all the assemblies within the instrument.

The query performs a full alignment and returns a number indicating the success of the alignment. A zero is returned if the alignment is successful. A one is returned if any part of the alignment failed.

Front Panel

Access: System, Alignments, Align All Now

Calibrate the Attenuator

:CALibration:ATTenuator

:CALibration:ATTenuator?

214