SERVice Subsystem

Language Reference

SERVice Subsystem

SERVice Subsystem

Provides SCPI access for the calibration manager.

Numeric values for bit patterns can be entered using decimal or hexidecimal representations. (i.e. 0 to 32767 is equivalent to #H0 to #H7FFF) See the SCPI Basics information about using bit patterns for variable parameters.

Prepare Calibration Files for Access

:SERVice[:PRODuction]:CALibrate:BEGin

Locks all of the calibration files for memory accesses.

Remarks:

No query.

Load Default Calibration Data to NRAM

:SERVice[:PRODuction]:CALibrate:DEFault <cal_fid>

Loads the specified calibration data from EEROM to NRAM, initializing the alignment data to the factory defaults..

Range:

cal_fid, corresponds to the Calibrate file ID

Remarks:

No query.

Unlock Calibration Files

:SERVice[:PRODuction]:CALibrate:END

Unlocks all of the calibration files.

Remarks: info

Store Calibration Data in EEROM

:SERVice[:PRODuction]:CALibrate:STORe <cal_fid>

Stores the specified calibration data into EEROM. The data will survive power cycles and will be reloaded into NRAM on power up.

Range:

cal_fid, corresponds to the calibration data file ID

Remarks:

No query.