Agilent Technologies E4406A VSA manual Questionable Event Query, STATus:QUEStionable:EVENt?

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 368
Image 368
Questionable Event Query

Language Reference

STATus Subsystem

 

Factory Preset

 

 

and *RST:

0

 

Range:

0 to 32767

 

Questionable Event Query

 

:STATus:QUEStionable[:EVENt]?

 

This query returns the decimal value of the sum of the bits in the

 

Questionable Event register.

 

The register requires that the associated PTR or NTR filters be set

NOTE

 

before a condition register bit can set a bit in the event register.

 

The data in this register is latched until it is queried. Once queried, the

 

register is cleared.

 

Key Type:

There is no equivalent front panel key.

Questionable Negative Transition

:STATus:QUEStionable:NTRansition <number>

:STATus:QUEStionable:NTRansition?

This command determines what bits in the Questionable Condition register will set the corresponding bit in the Questionable Event register when the condition register bit has a negative transition (1 to 0). The variable <number> is the sum of the decimal values of the bits that you want to enable.

Key Type:

There is no equivalent front panel key.

Factory Preset

 

and *RST:

0

Range:

0 to 32767

Questionable Positive Transition

:STATus:QUEStionable:PTRansition <number>

:STATus:QUEStionable:PTRansition?

This command determines what bits in the Questionable Condition register will set the corresponding bit in the Questionable Event register when the condition register bit has a positive transition (0 to 1). The variable <number> is the sum of the decimal values of the bits that you want to enable.

368

Chapter 5

Page 368
Image 368