Adjacent Channel Power—Offset Frequency Power Mode

Language Reference

SENSe Subsystem

Adjacent Channel Power—Offset Frequency Power Mode

[:SENSe]:ACP:LIST:POWer

INTegPEAK,INTegPEAK,INTegPEAK,INTegPEAK,INTegPEAK

[:SENSe]:ACP:LIST:POWer?

Define the power measurement mode for each of the offset frequencies. You can turn off (not use) specific offsets with the SENS:ACP:LIST:STATe command.

Factory Preset

 

and *RST:

INTeg, INTeg, INTeg, INTeg, INTeg

Remarks:

You must be in the NADC mode to use this command.

 

Use INSTrument:SELect to set the mode.

Adjacent Channel Power—Offset Frequency Relative Limit

[:SENSe]:ACP:LIST:RLIMit

<rel_powr>,<rel_powr>,<rel_powr>,<rel_powr>,<rel_powr>

[:SENSe]:ACP:LIST:RLIMit?

Set the relative limit on offset frequencies. You can turn off (not use) specific offsets with the SENS:ACP:LIST:STATe command.

Factory Preset

 

 

 

 

and *RST:

45 dB

 

 

 

 

 

 

 

 

 

 

Mode

Offset A

Offset B

Offset C

Offset D

Offset E

 

 

 

 

 

 

NADC

26 dB

45 dB

45 dB

0 dB

0 dB

 

 

 

 

 

 

PDC

45 dB

60 dB

0 dB

0 dB

0 dB

 

 

 

 

 

 

Range:

200 to 50 dB

 

 

 

Remarks:

You must be in the NADC, cdmaOne, or PDC mode to

 

use this command. Use INSTrument:SELect to set the

 

mode.

 

 

 

 

Adjacent Channel Power—Offset Frequency Control

[:SENSe]:ACP:LIST:STATe OFFON01, OFFON01, OFFON01, OFFON01, OFFON01

[:SENSe]:ACP:LIST:STATe?

Turn measurement on or off for the custom offset frequencies.