Agilent Technologies E4406A VSA manual Spectrum - Y-AxisReference Level, <power>

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 231
Image 231
Spectrum - Y-Axis Reference Level

Language Reference

DISPlay Subsystem

n=2, m=1 I Waveform (Option B7C)

n=2, m=2 Q Waveform (Option B7C)

n=3, m=1 I/Q Polar (Basic, W-CDMA, cdma2000)

n=4, m=1 Linear Spectrum (Basic, W-CDMA, cdma2000) m selects the window within the view. The default is 1.

Factory Preset: 10 dB per division, for Spectrum

Range:

.1 dB to 20 dB per division, for Spectrum

Default Unit:

10 dB per division, for Spectrum

Remarks:

May affect input attenuator setting.

 

To use this command, the appropriate mode should be

 

selected with INSTrument:SELect.

Front Panel

 

Access:

When in Spectrum measurement: Amplitude Y Scale,

 

Scale/Div.

History:

Modified revision A.05.00

Spectrum - Y-Axis Reference Level

:DISPlay:SPECtrum[n]:WINDow[m]:TRACe:Y[:SCALe]:RLEVel

<power>

:DISPlay:SPECtrum[n]:WINDow[m]:TRACe:Y[:SCALe]:RLEVel?

Sets the amplitude reference level for the y-axis. n selects the view, the default is Spectrum.

n=1, m=1 Spectrum

n=1, m=2 I/Q Waveform

n=1, m=2 I and Q Waveform (Basic, W-CDMA, cdma2000)

n=1, m=3 numeric data (Service mode)

n=1, m=4 RF envelope (Service mode)

n=2, m=1 I Waveform (Option B7C)

n=2, m=2 Q Waveform (Option B7C)

n=3, m=1 I/Q Polar (Basic, W-CDMA, cdma2000)

n=4, m=1 Linear Spectrum (Basic, W-CDMA, cdma2000) m selects the window within the view. The default is 1.

Factory Preset: 0 dBm, for Spectrum

Range: 250 to 250 dBm, for Spectrum

Chapter 5

231

Page 231
Image 231
Agilent Technologies E4406A VSA Spectrum - Y-AxisReference Level, DISPlay:SPECtrumn:WINDowm:TRACe:Y:SCALe:RLEVel, <power>