Agilent Technologies E4406A VSA manual Language Reference, MEASure Group of Commands

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 262
Image 262
Manual background

Language Reference

MEASure Group of Commands

Measurement

n

Results Returned

Type

 

 

 

 

 

 

 

3

Returns 10 comma-separated scalar values of the pass/fail

 

NADC and

(0=passed, or 1=failed) results determined by testing the

 

relative power of the offset frequencies:

 

PDC mode

 

 

 

 

 

1.

Negative offset frequency (1) relative power

 

 

2.

Positive offset frequency (1) relative power

 

 

 

. . .

 

 

1.

Negative offset frequency (5) relative power

 

 

2.

Positive offset frequency (5) relative power

 

 

 

 

3

Returns 3 comma-separated scalar values of the histogram

 

iDEN

relative power trace:

 

 

 

 

mode

1.

Lower offset frequency relative power

 

 

2.

Reference frequency relative power

 

 

3.

Upper offset frequency relative power

 

 

 

Power spectral

3

Returns 11 comma-separated scalar values (in dBm/Hz)

density

Basic,

corresponding to the power spectral density histogram display.

reference

The values are returned in ascending frequency order:

cdmaOne,

 

 

 

 

cdma2000,

1.

Negative offset frequency (5)

 

W-CDMA

2.

Negative offset frequency (4)

 

(3GPP), or

 

. . .

 

W-CDMA

 

 

(Trial &

1.

Center frequency

 

Arib) mode

 

2.

Positive offset frequency (1)

 

 

 

 

 

. . .

 

 

1.

Positive offset frequency (5)

 

 

 

 

4

Returns the frequency-domain spectrum trace (data array) for

 

NADC and

the entire frequency range being measured.

 

 

 

 

PDC mode

In order to return spectrum data, the ACP display must be in

 

 

the spectrum view and you must not turn off the spectrum

 

 

trace.

 

 

 

 

4

Returns 4 comma-separated absolute power results for the

 

iDEN

reference and offset channels.

 

 

 

 

mode

1.

Reference channel absolute power

 

 

2.

Reference channel absolute power (duplicate of above)

 

 

3.

Lower offset channel absolute power

 

 

4.

Upper offset channel absolute power

 

 

 

 

262

Chapter 5

Page 262
Image 262
Agilent Technologies E4406A VSA manual Language Reference, MEASure Group of Commands