Agilent Technologies E4406A VSA manual Power vs. Time—Numberof Bursts Averaged

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 335
Image 335
Power vs. Time—Number of Bursts Averaged

 

Language Reference

 

SENSe Subsystem

Remarks:

You must be in the Basic, cdma2000, W-CDMA (3GPP),

 

or W-CDMA (Trial & Arib) mode to use this command.

 

Use INSTrument:SELect to set the mode.

Power vs. Time Measurement

Commands for querying the power versus time measurement results and for setting to the default values are found in the “MEASure Group of Commands” on page 255. The equivalent front panel keys for the parameters described in the following commands, are found under the Meas Setup key, after the Power vs Time measurement has been selected from the MEASURE key menu.

Power vs. Time—Number of Bursts Averaged

[:SENSe]:PVTime:AVERage:COUNt <integer>

[:SENSe]:PVTime:AVERage:COUNt?

Set the number of bursts that will be averaged. After the specified number of bursts (average counts), the averaging mode (terminal control) setting determines the averaging action.

Factory Preset

 

and *RST:

15

Range:

1 to 10,000

Remarks:

You must be in the EDGE(w/GSM), GSM or Service

 

mode to use this command. Use INSTrument:SELect to

 

set the mode.

Power vs. Time—Averaging State

[:SENSe]:PVTime:AVERage[:STATe] OFFON01

[:SENSe]:PVTime:AVERage[:STATe]?

Turn averaging on or off.

Factory Preset

 

and *RST:

OFF

Remarks:

You must be in the EDGE(w/GSM), GSM or Service

 

mode to use this command. Use INSTrument:SELect to

 

set the mode.