MEASure Group of Commands

 

 

n

Results Returned

 

 

8, only available

Returns comma-separated trace points of the minimum waveform data.

when averaging

These data points are floating point numbers representing the power of the

is set to both

signal (in dBm). There are N data points, where N is the number of

maximum and

samples. The period between the samples is defined by the sample time.

minimum

Use SENSe:PVT:AVERage:TYPE MXMinimum to set averaging to max and

 

 

min. Use n=2 to return the corresponding maximum trace.

 

 