Agilent Technologies E4406A VSA SENSe:ACP:OFFSet:LIST:RPSDensity?, cdma2000, W-CDMA3GPP mode

Models: E4406A VSA

1 406
Download 406 pages 63.49 Kb
Page 306
Image 306
[:SENSe]:ACP:OFFSet:LIST:RPSDensity?

Language Reference

SENSe Subsystem

[:SENSe]:ACP:OFFSet:LIST:RPSDensity?

cdma2000, W-CDMA (3GPP) mode

[:SENSe]:ACP:OFFSet[n]:LIST:RPSDensity

<rel_power>,<rel_power>,<rel_power>,<rel_power>,<rel_power>

[:SENSe]:ACP:OFFSet[n]:LIST:RPSDensity?

cdmaOne, W-CDMA (Trial & Arib) mode

[:SENSe]:ACP:OFFSet[n]:LIST[n]:RPSDensity

<rel_power>,<rel_power>,<rel_power>,<rel_power>,<rel_power>

[:SENSe]:ACP:OFFSet[n]:LIST[n]:RPSDensity?

Sets the amplitude levels to test against for any custom offsets. This amplitude level is relative to the power spectral density. If multiple offsets are available, the list contains five (5) entries. The offset closest to the carrier channel is the first one in the list. [:SENSe]:ACP:OFFSet[n]:LIST[n]:TEST selects the type of testing to be done at each offset.

You can turn off (not use) specific offsets with the [:SENSe]:ACP:OFFSet[n]:LIST:STATe command.

The query returns five (5) real numbers that are the current amplitude test limits, relative to the power spectral density, for each offset.

Offset[n]

n=1 is base station and 2 is mobiles. The default is base

 

station (1).

List[n]

 

cdmaOne mode n=1 is cellular bands and 2 is pcs bands. The default is cellular.

W-CDMA (Trial

&Arib) mode n=1 is ARIB, 2 is 3GPP, and 3 is Trial. The default is ARIB (1).

Factory Preset and *RST:

Mode

Variant

Offset A

Offset B

Offset C

Offset D

Offset E

 

 

 

 

 

 

 

iDEN

 

0 dB

n/a

n/a

n/a

n/a

 

 

 

 

 

 

 

Basic

 

28.87 dB

43.87 dB

0 dB

0 dB

0 dB

 

 

 

 

 

 

 

cdmaOne

BS cellular

28.87 dB

43.87 dB

0 dB

0 dB

0 dB

 

 

 

 

 

 

 

 

BS pcs

28.87 dB

0 dB

0 dB

0 dB

0 dB

 

 

 

 

 

 

 

 

MS cellular

25.87 dB

37.87 dB

0 dB

0 dB

0 dB

 

 

 

 

 

 

 

 

MS pcs

25.87 dB

0 dB

0 dB

0 dB

0 dB

 

 

 

 

 

 

 

306

Chapter 5

Page 306
Image 306
Agilent Technologies E4406A VSA manual SENSe:ACP:OFFSet:LIST:RPSDensity?, SENSe:ACP:OFFSetn:LIST:RPSDensity, W-CDMATrial